Belated evaluation and reporting of this complaint was done during retrospective review as part of capa 20-0052 corrective action 6.The 8401hx c-mac® video laryngoscope d-blade was sent to karl storz on 23rd december 2020 for investigation.Upon investigation it can be confirmed that the c mac blade was defective.The covering glass and the lens are missing.The bad image quality, as reported, during the procedure is most likely caused by a defective camera and/or residues present at the interfaces.As per confirmation by the hospital, the d-blade used was defective before the procedure, however the user/nurses did not remove the defective blade from the holder and also the user did not check the function of the device before use.As a consequence, the defective device was then used and caused a bad quality image during the procedure.The root cause is most likely due to user error to use the defective device and not check its function prior to its use.In the corresponding ifu 96076011en-us, v3.1 11/2019 below chapter 5.1 the user must to perform a visual and functional test before and after every use.Risk level and probability are still adequate.The event is filed under internal karl storz complaint id (200843848).
|