SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VH
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus and the customer¿s complaint was confirmed.Additional evaluation findings are as follows: 1.) bending section cover noted leaking; 2.) distal cover insulation measured at 10 megaohms; 3.) multiple components scratched; 4.) multiple non-olympus parts noted; 5.) control knob markings almost illegible; 6.) rubber around the mouthpiece was noted detached.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus that during reprocessing they discovered that the broken rubber o-ring was broken causing a gap in the instrument port.There were no reports of patient harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It was confirmed that the o-ring rubber was broken but the cause could not be specified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16950753
MDR Text Key315461079
Report Number3002808148-2023-04959
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170310461
UDI-Public04953170310461
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K221683
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/17/2023
Initial Date FDA Received05/17/2023
Supplement Dates Manufacturer Received06/02/2023
Supplement Dates FDA Received06/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-