Model Number 1944/52
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/08/2023
Event Type  Injury  
Event Description
It was reported the patient present in post-implant checks.Interrogation revealed the atrial lead failed to sense p-waves.The atrial lead was explanted and replaced.The patient was in stable condition.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The reported event of failure to sense was not confirmed.A complete lead was received for analysis.Electrical and x-ray evaluations were normal with no anomalies found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16955083
MDR Text Key315488918
Report Number2017865-2023-20229
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734502320
UDI-Public05414734502320
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1944/52
Device Catalogue Number1944-52
Device Lot NumberP000154797
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/08/2023
Initial Date FDA Received05/18/2023
Supplement Dates Manufacturer Received07/10/2023
07/13/2023
Supplement Dates FDA Received07/13/2023
07/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-