Model Number 16-02-85
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2023
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in hudson, florida.A livanova field service representative was dispatched to the facility to investigate the device and could not confirm the reported issue, since the device was emptied by the customer.Level sensor shafts were cleaned by livanova technician in case these were sticking preventing the customer from knowing when the patient tank was full.Missing bracket for the vacuum canister was installed.Unit was filled with water and tested: all tests passed showing no signs of leaks.The unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t device was leaking water from front and back, during cleaning operations.There was no patient involvement.
 
Manufacturer Narrative
H10: based on available information, it cannot be ruled out that leak was caused by the user who inadvertently overfilled the unit during cleaning operation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16959835
MDR Text Key315570687
Report Number9611109-2023-00226
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/21/2023
Initial Date FDA Received05/19/2023
Supplement Dates Manufacturer Received06/29/2023
Supplement Dates FDA Received07/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-