Model Number BF-Q190 |
Device Problems
Device Reprocessing Problem (1091); Microbial Contamination of Device (2303)
|
Patient Problem
Fungal Infection (2419)
|
Event Date 02/03/2021 |
Event Type
Injury
|
Event Description
|
The customer reported to olympus, the evis exera bronchovideoscope bronch wash tested positive for sarocladium during a routine culture of the scope.The sampling was taken at reprocessing, before use.Additional information received indicated there were 16 patients with positive culture results between march 20, 2018 and february 25, 2022.None of the patients were treated for this organism.The following medwatch reports are related: complaint numbers model#/ serial # procedure date patient 1 - (b)(6) bf-q190 / (b)(6) (b)(6) 2019.Patient 1 - (b)(6) bf-uc180f / (b)(6) (b)(6) 2019.Patient 2 - (b)(6) bf-q190 / (b)(6) (b)(6) 2020.Patient 2 - (b)(6) bf-uc180f / (b)(6) (b)(6) 2020.Patient 3 - (b)(6) bf-1t180 / (b)(6) admit: (b)(6) 2020, bronch: (b)(6) 2020.Patient 4 - (b)(6) bf-q190 / (b)(6) (b)(6) 2020.Patient 5 - (b)(6) bf-uc180f / (b)(6) (b)(6) 2020.Patient 6 - (b)(6) bf-1th190 / (b)(6) (b)(6) 2020.Patient 6 - (b)(6) bf-uc180f / (b)(6) (b)(6) 2020.Patient 7 - (b)(6) bf-uc180f / (b)(6) (b)(6) 2020.Patient 8 - (b)(6) bf-1th190 / (b)(6) (b)(6) 2020.Patient 8 - (b)(6) bf-uc180f / (b)(6) (b)(6) 2020.Patient 9 - (b)(6) bf-q190 / (b)(6) (b)(6) 2020.Patient 10 - (b)(6) bf-1th190 / (b)(6) (b)(6) 2021.Patient 10 - (b)(6) bf-uc180f / (b)(6) (b)(6) 2021.Patient 11 - (b)(6) bf-q190 / (b)(6) (b)(6) 2021.Patient 12- (b)(6) bf-1th190 / (b)(6) (b)(6) 2021.Patient 12 - (b)(6) bf-uc180f / (b)(6) (b)(6) 2021.Patient 13 - (b)(6) bf-1t180 / (b)(6) (b)(6) 2021.Patient 13 - (b)(6) bf-uc180f / (b)(6) (b)(6) 2021.Patient 14 - (b)(6) bf-q190 / (b)(6) (b)(6) 2021.Patient 15 - (b)(6) bf-p190 / (b)(6) (b)(6) 2021.Patient 16 - (b)(6) bf-q190 / (b)(6) (b)(6) 2022.This medwatch report is for patient identifier (b)(6): infection verified via bronchoalveolar lavage, results (b)(6) 2021talaromyces sp not t.Marneffei, sarocladium species.Patient's current condition alive.
|
|
Manufacturer Narrative
|
The events were discovered with the culture results as identified for the patient listed.The user facility provided additional information regarding the cleaning, the disinfection and the sterilization processes performed onsite for the endoscopes.Accordingly, to the customer the scope was reprocessed after every use with no delay, and water was aspirated through the instrument/suction channel with a suction pump.No abnormalities were found with the reprocessing accessories and the scope passed leak testing.Manual cleaning was also performed 1 hr.After the procedure and the scope instrument/suction/balloon channel, air/water/suction/biopsy valve were brushed, and pure (boston scientific) detergent solution was used.For automated endoscope reprocessor (aer) treatment, the medivators reprocessor along with intercept plus detergent and rapacide pa disinfectant were used.The reprocessor and the scope were not tested.As part of their own investigation, the facility invited state surveyors to come and investigate their processes which did not turn up any definitive conclusions as to the cause of these positive cultures.The customer decided to send one scope to olympus ((b)(6) bf-q190 / (b)( 6)).Depending on findings, the facility will discuss and decide on the need to send the additional scopes in for investigation.The scope was sent to nelson lab by olympus for independent culture testing.The customer reported that the last reprocessing in-service conducted by an olympus endoscopy support specialist was jun 08, 2022.An olympus endoscopy support specialist (ess) visit was scheduled for a future date to observe the user facility¿s reprocessing practices from start to finish and provide reprocessing in-service training.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.This report has been submitted by the importer under this mdr report number 2429304 -2023-00136.
|
|
Event Description
|
Additional information received indicating all the reported scopes were utilized on patents who had positive cultures for sarocladium post bronchoscopy procedure.The facility has not identified sarocladium in the scopes themselves.All of these cases were determined by dr.Tjaden to not actually be infected as no signs or symptoms correlated with a fungal infection, no treatment or medical intervention done.There were no nosocomial infection cases in the hospital, involving the same microorganisms detected in the reported patients, during the period between 2019 and 2022.There no evidence found of contamination of bronchoalveolar lavages during collection and/or microbiological culture at this time.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information from the customer.New information added to the following field: b5.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information from the customer and to provide additional information based on the legal manufacturer's final investigation.Please note that the device sn (b)(6) was returned, and an evaluation/testing of the device was documented under related record patient identifier # (b)(6).A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the investigation, the root cause of the reported event could not be identified.Per the olympus hygiene microbiological investigation (hmi), the scope was negative for the reported microorganism (sarocladium fungi), but positive for a different microorganism (staphylococcus epidermis).However, obvious deviation from the instruction for use (ifu) regarding device reprocessing was not observed (per the information provided by the facility).Furthermore, in this case there was a report of 16 infected patients with 14 of them positive for sarocladium fungi (per the bronchoalveolar lavage (bal) samples).There was a total of 6 different olympus bf-endoscopes used.Bf endoscopes varied in series (190 and 180) but also in model (p/q/1t/ebus).Although it is possible that 6 different bf endoscopes may have caused the patient infections; there are many other possible causes which could have led to the infections.The event can be detected/prevented by following the ifu which states: ¿an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them.¿ olympus will continue to monitor the field performance of this device.
|
|