SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/01/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and customer allegation was confirmed.The bending section cover exploded/removed, and the repair center was unable to perform leak test.The adhesive of the bending section cover was lifted/chipped and the image center was off.The red paint on the control body was missing.The investigation is ongoing.A supplemental will be submitted upon completion of investigation or if any additional information is received.
 
Event Description
The customer reported to olympus, the distal end cover of the oes cystonephrofiberscope was "blown off" and visibly broken.The issue was found during reprocessing.There were no reports of patient harm associated with the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, the event likely occurred due to use stress, external factors, or handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16973179
MDR Text Key316389550
Report Number3002808148-2023-05095
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339417
UDI-Public04953170339417
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K221690
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/02/2023
Initial Date FDA Received05/22/2023
Supplement Dates Manufacturer Received06/07/2023
Supplement Dates FDA Received07/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/05/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-