MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
|
Back to Search Results |
|
Model Number BI70000028120 |
Device Problems
Failure to Run on Battery (1466); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/18/2023 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that an issue with mobile view station (mvs) was still occurring.When the site would unplug the system in the hallway to bring into the room, it wouldn't hold a charge and would die in 30 seconds.Every time they unplugged the mvs, it died within the 30 second time frame.No patient was present.
|
|
Manufacturer Narrative
|
No parts have been received by the manufacturer for evaluation.Codes b17, c20, d15 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H2) device evaluation: h3, h6: a medtronic representative went to the site to test the equipment.Testing revealed that the manufacturer representative unp lugged the mobile viewing system (mvs) from wall power and started the stopwatch to find that the mvs stayed on for over five minutes as the previous visit.The uninterruptible power supply (ups) functioned properly at the time.System checkout was performed and the system returned to service.B01, c19, d14 are applicable.H2) correction: additional code img code updated from g02001 to g07001.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|