• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC MOD EX 5.5 QC SCREWDRIVER; SREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC MOD EX 5.5 QC SCREWDRIVER; SREWDRIVER Back to Search Results
Catalog Number 698343530
Device Problems Break (1069); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Event Description
Tip of screwdriver is broken on two screwdrivers during insertion of screws.Happened at two different surgeries and patients.Not at the same time tip of screwdriver was fixed in the screw.Screw was removed and new screw was placed with another screwdriver.No consequence for patient.This report is for one (1) mod ex 5.5 qc screwdriver.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This report is related to (b)(4), which captures the additional screwdriver and surgery mentioned.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6 investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that mod ex 5.5 qc screwdriver was broken at the hex star tip.The broken fragment was not returned.The reported complaint condition was able to be confirmed.A dimensional inspection was unable to be performed due to post-manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the mod ex 5.5 qc screwdriver would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: yes, reviewed.Dimensional inspection: n/a.Device history: a manufacturing related potential cause was not suspected, therefore, per franchise complaint product investigation procedure no manufacturing record evaluation is required.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOD EX 5.5 QC SCREWDRIVER
Type of Device
SREWDRIVER
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16974964
MDR Text Key315681643
Report Number1526439-2023-00884
Device Sequence Number1
Product Code HXX
UDI-Device Identifier10705034534091
UDI-Public(01)10705034534091
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number698343530
Device Lot NumberPRODUCT CODE WRITTEN ON INSTRU
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/05/2023
Initial Date FDA Received05/22/2023
Supplement Dates Manufacturer Received05/22/2023
07/21/2023
Supplement Dates FDA Received06/06/2023
08/03/2023
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-