There was an allegation of questionable elecsys troponin t hs assay results for 5 patient samples on 3 cobas e 801 analytical units with serial numbers (b)(6), (b)(6), and (b)(6).The initial results were reported outside of the laboratory.The customer was performing all troponin testing in duplicate due to previous issues.For sample 1, the initial troponin result was 15.2 pg/ml on analyzer serial number (sn) (b)(6).The repeat result was 21.5 pg/ml on analyzer sn (b)(6).For sample 2, the initial troponin result was 759 pg/ml on analyzer sn (b)(6).The repeat result was 944 pg/ml on analyzer sn (b)(6).For sample 3, the initial troponin result was 593 pg/ml on analyzer sn (b)(6).The repeat result was 733 pg/ml on analyzer sn (b)(6).For sample 4, the initial troponin result was 654 pg/ml on analyzer sn (b)(6).The repeat result was 799 pg/ml on analyzer sn (b)(6).For sample 5, the initial troponin result was 22.4 pg/ml on analyzer sn (b)(6).The repeat result was 27.1 pg/ml on analyzer sn (b)(6).
|
Calibration was last performed on (b)(6) 2023.The qc recovery data provided was acceptable.Based on the calibration and qc data, a general reagent issue could be excluded.The alarm trace showed several sample clot, sample short, sample foam, and abnormal aspiration alarms across all three analyzers.The investigation did not identify a product problem.The root cause of the event could not be determined.
|