Model Number 33600020
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2023
Event Type  malfunction  
Event Description
The holes in the guide arm were to small to properly accommodate pin sleeves.The pin sleeves needed to be hammered in and, after the procedure, hammered out.
 
Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALIGNMENT FRAME DIST SUB ASSEM INFINITY TOTAL ANKLE SYSTEM
Type of Device
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16979872
MDR Text Key315759934
Report Number3010667733-2023-00268
Device Sequence Number1
Product Code HSN
UDI-Device Identifier00889797003926
UDI-Public00889797003926
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K140749
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number33600020
Device Catalogue Number33600020
Device Lot NumberNOT REPORTED
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/24/2023
Initial Date FDA Received05/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse