Model Number 33600020
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2023
Event Type  malfunction  
Manufacturer Narrative
The reported event could be confirmed since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: visual examination of the returned device shows no visible deformation around the instrument.A functional test was performed using a pin sleeve with the same part number and a different lot number as those used in the reported event.The pin sleeve was able to pass through the three proximal holes of the device but failed to fit through the three distal holes as reported.Gage pins were used to verify the dimensions of the distal holes of the device.A 0.232862 minus pin and a 0.232519 plus pin were used to perform the dimensional inspection.Both pins were unable to fit through the distal holes of the device.Therefore, the distal holes of the device were confirmed to be out of specification.Based on investigation, the root cause was attributed to a manufacturing/supplier related issue.A nonconformance was opened to further investigate the issue.
 
Event Description
It was reported that the sleeve did not pass through the guide arm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALIGNMENT FRAME DIST SUB ASSEM INFINITY TOTAL ANKLE SYSTEM
Type of Device
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16979891
MDR Text Key315797257
Report Number3010667733-2023-00269
Device Sequence Number1
Product Code HSN
UDI-Device Identifier00889797003926
UDI-Public00889797003926
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K140749
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 05/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number33600020
Device Catalogue Number33600020
Device Lot Number2656815
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2023
Initial Date Manufacturer Received 04/24/2023
Initial Date FDA Received05/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-