WRIGHT MEDICAL TECHNOLOGY INC ALIGNMENT FRAME DIST SUB ASSEM INFINITY TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 33600020 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported event could be confirmed since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: visual examination of the returned device shows no visible deformation around the instrument.A functional test was performed using a pin sleeve with the same part number and a different lot number as those used in the reported event.The pin sleeve was able to pass through the three proximal holes of the device but failed to fit through the three distal holes as reported.Gage pins were used to verify the dimensions of the distal holes of the device.A 0.232862 minus pin and a 0.232519 plus pin were used to perform the dimensional inspection.Both pins were unable to fit through the distal holes of the device.Therefore, the distal holes of the device were confirmed to be out of specification.Based on investigation, the root cause was attributed to a manufacturing/supplier related issue.A nonconformance was opened to further investigate the issue.
|
|
Event Description
|
It was reported that the sleeve did not pass through the guide arm.
|
|