PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD20 BIPLANE
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2022
Event Type  malfunction  
Event Description
It has been reported to philips that an 'insufficient disk space' message appeared and there were almost no examinations in memory.The system was in clinical use when this occurred.There was no report of harm.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.The philips remote service engineer (rse) inspected the system remotely and confirmed that the image disks were almost full.The rse examined the system and found that the issue was with ippc (image processing pc) in the m cabinet.The rse recreated the image store of the frontal and lateral ippc to resolve the reported issue.After the repair, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16982019
MDR Text Key315740939
Report Number3014128039-2023-00001
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059061
UDI-Public00884838059061
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20 BIPLANE
Device Catalogue Number722013
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/05/2022
Initial Date FDA Received05/23/2023
Supplement Dates Manufacturer Received10/12/2023
Supplement Dates FDA Received11/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-