Model Number ALLURA XPER FD20
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2023
Event Type  malfunction  
Event Description
It has been reported to philips that an 'image disk full' message appeared and the procedure could not be continued.The system was in clinical use when this occurred.There was no report of harm.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed the system was not getting on.Review of the log files showed malfunction in the frontal ip-pc (image processing-pc).The fse examined the system and found that there was malfunction with the sectors of the hard drives as it was damaged.The fse recreated the image store but it didn't resolve the issue.The fse further investigated and found that there was issue with hard drives.The fse replaced the three hard drives in the ippc and loaded its software.After the replacement, the system was returned to use in good working order.The codes were updated based on the investigation outcome.Evaluation method code was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16985126
MDR Text Key315770920
Report Number3003768277-2023-03082
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059054
UDI-Public00884838059054
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20
Device Catalogue Number722012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/08/2023
Initial Date FDA Received05/23/2023
Supplement Dates Manufacturer Received11/11/2023
Supplement Dates FDA Received11/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-