MAUDE Adverse Event Report: EDWARDS LIFESCIENCES PR SWAN GANZ VIP OXIMETRY THERMOD; CATHETER, OXIMETER, FIBER OPTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES PR SWAN GANZ VIP OXIMETRY THERMOD; CATHETER, OXIMETER, FIBER OPTIC Back to Search Results
Model Number 782F75M
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2023
Event Type  malfunction  
Event Description
It was reported that the cardiac index obtained by thermodilution method did not match values compared to fick method.An edwards lifesciences 7.5 fr swan ganz was used during the incident.There were no patient injuries and no troubleshooting steps and additional procedures were noted.
 
Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation.The lot number was not provided thus a device history record was not reviewed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Manufacturer Narrative
The reported event of cardiac index issue was unable to be confirmed.The catheter was submerged in a 37.0c water bath and read 37.0c on vigilance ii monitor.Thermistor temperature reading accuracy is plus negative 0.3c per vigilance ii manual.Thermistor circuit was continuous.There were no open or intermittent conditions.Thermistor connector was opened and found no visible inconsistencies.Balloon inflated clear and concentric and remained inflated for 5 minutes without leakage.All through lumens were patent without any leakage or occlusion.No visible damage was observed from catheter body or returned syringe.Further evaluation regarding related quality issues is under investigation.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN GANZ VIP OXIMETRY THERMOD
Type of Device
CATHETER, OXIMETER, FIBER OPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
MDR Report Key16996287
MDR Text Key315915212
Report Number2015691-2023-13210
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K926450
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number782F75M
Device Catalogue Number782F75M
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/04/2023
Initial Date FDA Received05/24/2023
Supplement Dates Manufacturer Received05/26/2023
Supplement Dates FDA Received06/08/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-