Model Number X |
Device Problems
Battery Problem (2885); Communication or Transmission Problem (2896); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/29/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Visual inspection was performed on the returned device.The reported battery problem was not confirmed.There were no issues noted on the transmitter component, and the battery voltage met the specification.Functional testing could not be performed due to the noted corrosion on the sensor chip assembly.The corrosion was likely due to liquid exposure during post-procedure handling.There was also a torn material noted on the distal tube, and the sensor chip assembly was also displaced from the distal tube.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product quality issue.The investigation was unable to determine a cause for the reported battery problem.It may be possible that a steady yellow light instead of a blinking yellow light was observed during the procedure which could lead to the signal calibration failure.A steady yellow light was likely caused by the torn distal tube which may cause inadvertent damage to the wire such open circuit or broken connection.Based on the received information, the cause of torn material and sensor chip assembly displacement could not be determined.It may be possible that there was an interaction between the pressurewire and hemostasis valve during insertion/withdrawal that could result in the noted torn material and sensor chip assembly displacement; however, this could not be confirmed as it was based on operational circumstances.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
Event Description
|
It was reported that the procedure was performed to treat a lesion in the mildly calcified left anterior descending (lad) artery.The pressurewire x wireless device was connected to the quantien system and inserted into the hemostasis valve; however, there was no signal.The pal lighting color pattern was blinking yellow.The device was unable to be calibrated or equalized and was not used.The procedure was continued with a second same device.There was no adverse patient effect and no clinically significant delay.Returned device analysis noted that the distal tube was torn, the sensor jacket was displaced, and corrosion was noted on the sensor chip assembly, distal tip coil and transition between the proximal and distal tube.Additional follow up with the site indicated that the site was unaware of the device damage, and it is not known when the damage occurred.There was no difficulty noted during withdrawal and no adverse patient effect.No additional information was provided.
|
|
Manufacturer Narrative
|
Visual inspection was performed on the returned device.The reported flashing yellow light was not confirmed.The investigation was unable to determine a cause for the reported flashing yellow light.It may be possible that a steady yellow light instead of a blinking yellow light was observed during the procedure which could lead to the reported issue.A steady yellow light was likely caused by the torn distal tube which may cause inadvertent damage to the wire such as an open circuit or broken connection.Based on the received information, the cause of torn material and sensor chip assembly displacement could not be determined.It may be possible that there was an interaction between the pressurewire and hemostasis valve during insertion/withdrawal that could result in the noted torn material and sensor chip assembly displacement; however, this could not be confirmed as it was based on operational circumstances.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.H6: medical device problem code 2885, - removed.
|
|
Search Alerts/Recalls
|
|