• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS Back to Search Results
Model Number 8606500
Device Problems Gas Output Problem (1266); Failure to Deliver (2338); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2023
Event Type  malfunction  
Event Description
Around 3:40 pm on (b)(6) 2023 machine went into vent fail.No injury reported.
 
Manufacturer Narrative
The investigation was just started.The result will be forwarded once the investigation was closed.H3 other text : on-going.
 
Event Description
Around 3:40 pm on (b)(6) 23 machine went into vent fail.No injury reported.
 
Manufacturer Narrative
The investigation was performed based on the device log file, the replaced motor and the returned pba vgc power.The parts were tested in the laboratory without finding any deviation from specification.According to the logfile analysis the device detected a too high and too fast pressure increase which finally resulted in a motor blockage and subsequently in an emergency shutdown of the ventilator.To prevent from damages, the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.Dräger finally concludes that the device responded as specified upon the detected situation with an autonomous shutdown while changing mode to man/spont (safety mode) accompanied by an audible and visible "ventilator fail" alarm.The logfile analysis has not shown any indications for a technical malfunction.It can be assumed that the unstable pressure situation was the result of a patient interaction as the event was during getting the patient off the ventilator.The device was tested and returned to use with no further problems reported.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key17000456
MDR Text Key316033687
Report Number9611500-2023-00164
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier04048675229674
UDI-Public(01)04048675229674(11)190813(17)240206(93)8606500-67
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8606500
Device Catalogue Number8606500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/11/2023
Initial Date FDA Received05/25/2023
Supplement Dates Manufacturer Received06/26/2023
Supplement Dates FDA Received07/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-