Model Number 60-02-50
Device Problems Infusion or Flow Problem (2964); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2023
Event Type  malfunction  
Manufacturer Narrative
H10: livanova deutschland manufactures the s5 system.The incident occurred in united states.A livanova intiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland has received a report some wrong issue with flow on drive of an s5 console during a procedure.Also of fan was making noise.There was no patient injury.
 
Manufacturer Narrative
H.10: through follow-up communication, livanova learned the involved centrifugal pump system with tubing clamp (scp) which was used during the surgery.Model/serial numbers have been updated to the dedicated d.4 section.In addition, it was learned that technical service requested has been canceled and therefore no livanova technical intervention is foreseen at the moment.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
A livanova field service engeneer contacted the biomed at the hospital and it was clarified the complained issue was a noise from the fan inside the control panel.This unit has reached its ¿end of life¿ so repairs cannot be completed.This is the reason customer needed to cancel the repair and the device was not troubleshooted.The event has been reassessed as not reportable since no device malfunction has occurred and the claimed noise has unlikely possibility to cause any patient injury.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCP CONTROL PANEL W/FAST CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17002169
MDR Text Key315928348
Report Number9611109-2023-00240
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 07/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-50
Device Catalogue Number60-02-50
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/27/2023
Initial Date FDA Received05/25/2023
Supplement Dates Manufacturer Received05/01/2023
06/15/2023
Supplement Dates FDA Received05/30/2023
07/11/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/23/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-