Catalog Number CLXECP
Device Problem Coagulation in Device or Device Ingredient (1096)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/25/2023
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.A batch record review for kit lot l121 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot l121 shows no trends.Trends were reviewed for complaint categories, alarm #17: return pressure and clot observed.No trends were detected for these complaint categories.This case is reportable as a mdr due to the reportable malfunction clot observed in the return line.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Section 2-9 of the cellex operators manual (1470493 rev 6) for use with software 5.4 on anticoagulant states "caution: individual patients may require a heparin dosage that varies from the recommended dose to prevent post-treatment bleeding or clotting during a treatment.The physician should review the patient's medical condition, medications and platelet count at the time of treatment and use clinical judgement to establish the optimal heparin dosage for each patient." the root cause for the clotting observed could not be determined based on the available information.No further action is required at this time.This investigation is now complete.(b)(4).N.S.25-may-2023.
 
Event Description
The customer contacted mallinckrodt to report they experienced clots observed in the return line with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported they received an alarm #17: return pressure alarm when approximately 600 ml of whole blood was processed.The customer paused the ecp treatment and noticed blood clotting in the return line.The ecp treatment was aborted and residual blood within the kit was not returned to the patient.There was no report of patient harm associated with this event.The customer did not return the product for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin, D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin, D15 T X2V
EI   D15 TX2V
Manufacturer Contact
mark wendelken
440 route 22 east
suite 302
bridgewater, NJ 08807
MDR Report Key17002460
MDR Text Key316385034
Report Number3013428851-2023-00026
Device Sequence Number1
Product Code LNR
Combination Product (y/n)Y
Reporter Country CodeSZ
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2024
Device Catalogue NumberCLXECP
Device Lot NumberL121
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/25/2023
Initial Date FDA Received05/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexFemale
Patient Weight64 KG
-
-