COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGTRSB60AMT
Device Problems Entrapment of Device (1212); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/02/2023
Event Type  malfunction  
Event Description
According to the reporter, during a laparotomy transverse colectomy, during side-to-side anastomosis, and when the insertion hole was closed, on two reloads, the removal of the anchoring suture at the tip was poor.There was also difficulty removing the reload from the cavity.The surgeon had to remove slightly stronger.There was no patient injury.
 
Manufacturer Narrative
D10 concomitant products: sigphandle, sig power sigphandle handle (sn:unknown) sigtrsb60amt, sigtrsb60amt 60mm med thk reinforced rel (lot#n2j0052y) unk-sigadapt, unknown signia egia adapter (sn:unknown) sigpshell, sig power sigpshell control shell (lot#unknown h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the reloads were fully fired and the interlocks were engaged, the jaws of the reloads were open, and the sutures were cut properly.Functionally, the reloads were loaded into a representative handle.The interlocks were overridden and the reloads were applied to test medias.Test medias were transected. the reload interlocks were tested and found to function properly.It was reported that the jaws of the device remain closed on tissue, and after firing, the reinforcement material did not release from the device as expected.The reported issues could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: the device had damage to the cutting edge of the knife blade.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: when positioning the stapler on the application site, ensure that no obstructions, such as clips, are incorporated into the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that jaws of the device remained closed on tissue.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17004439
MDR Text Key315972258
Report Number1219930-2023-02112
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884521717626
UDI-Public10884521717626
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173270
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSIGTRSB60AMT
Device Catalogue NumberSIGTRSB60AMT
Device Lot NumberN2J0052Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/09/2023
Initial Date FDA Received05/25/2023
Supplement Dates Manufacturer Received09/01/2023
Supplement Dates FDA Received09/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-