Patient was placed on table for an arthrogram.Before the injection, the fluoro monitor in the room went black.The fluoro monitor in the control area still worked, however a re-boot of the system did not resolve the issue.Patient was moved to a different room and procedure was completed.No harm to the patient except delay in case.Room had a fco completed on it the day before - (b)(4) under philips wo (b)(4).Room was tested after the fco was completed and again the morning of the case, but when this first case was started, system failed.Unsure if this is related to the fco or the work done by philips technician (b)(4).Philips was called back in to review and resolve.Philips calibrated the monitor.Adjusted exam room monitor vga to hdmi then all monitors are functional.Tested the system returned to clinical use.Philips representative (b)(4) indicated he would follow up with nss (b)(4) to see if this calibration of the exam room monitor that muzapaer completed this morning after the room was in use, should have been done during the fco.According to (b)(4), either the fco procedure is lacking or (b)(4) missed steps.Either situation is not acceptable and philips needs to prevent this for future fco's/repairs.
|