• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SETS - FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SETS - FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7346-24
Device Problems Device Alarm System (1012); Partial Blockage (1065); Complete Blockage (1094); Obstruction of Flow (2423); Blocked Connection (2888)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/24/2023
Event Type  malfunction  
Manufacturer Narrative
Other, other text: a supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported the disposable made the device alarm "down stream occlusion".Patient was sent to fmc hptp to have additional troubleshooting done and the remainder of vancomycin infused.No patient injury reported.
 
Manufacturer Narrative
One disposable was received for evaluation.Visual inspection found the disposable to be in used condition, and the bond between the coil tube and entrance of the filter was observed to be occluded.To conduct functional testing, the disposable was primed and set to be ran with a cadd legacy pump.He fluid did not pass through the unit and the alarm high pressure? was activated.The reported failure mode, pump alarm, is confirmed.The root cause was determined to be the occlusion in the bond between the coil tube and filter entrance due to an excess of solvent.This was caused by failure to properly follow manufacturing procedure.No lot number was provided; therefore, a device history record (dhr) review could not be conducted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SETS - FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17010454
MDR Text Key316039420
Report Number3012307300-2023-05736
Device Sequence Number1
Product Code FPA
UDI-Device Identifier15019517161953
UDI-Public15019517161953
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K031361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7346-24
Device Catalogue Number21-7346-24
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/09/2023
Initial Date FDA Received05/26/2023
Supplement Dates Manufacturer Received07/28/2023
Supplement Dates FDA Received10/16/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexFemale
-
-