U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST; FORCE BIPOLAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST; FORCE BIPOLAR Back to Search Results
Model Number 470405-06
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/30/2020
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis (fa) investigation confirmed that the instrument was driven and moved intuitively with full range of motion in all directions.No difficulty removing and installing the instrument on the system arm.Levers were still present on the instrument and were easily depressed to remove the instrument during disengagement of instrument with various cannulas.
 
Event Description
It was reported that during a da vinci-assisted total benign hysterectomy, the customer had issues with the force bipolar (fb) instrument removal.The customer had to remove the trocar and use another instrument.The procedure was completed with no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
FORCE BIPOLAR
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key17015988
MDR Text Key316159639
Report Number2955842-2023-15098
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874115930
UDI-Public(01)00886874115930(10)N10191216
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180351
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470405-06
Device Catalogue Number470405
Device Lot NumberN10191216 0035
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2020
Initial Date Manufacturer Received 06/30/2020
Initial Date FDA Received05/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-