Intuitive surgical, inc.(isi) received the force bipolar (fb) instrument involved with this complaint and completed the device evaluation.Failure analysis (fa) investigation did not replicate nor confirm the customer reported complaint of breakage of the cable.The customer also reported a complaint of a "bent jaw." this complaint was confirmed.The instrument was found to have a severely bent grip, and the tip did not align with the other grip.This failure is most commonly caused by mishandling/misuse, such as excess force applied to the instrument jaws.
|
It was reported that during a da vinci-assisted inguinal hernia repair surgical procedure, the surgeon had a collision with the monopolar curved scissors (mcs) instrument and a force bipolar (fb) instrument.As a result, the fb instrument's cable was broken and the jaws were bent.An instrument release kit (irk) was used to remove the instrument from the cannula.The procedure was completed with no report of patient harm, injury or adverse outcome.Intuitive surgical, inc.(isi) follow-up with the customer and confirmed no fragment fell inside the patient.The surgeon believes the cause of the instrument breakage was from the arm collision.The instrument was reportedly inspected prior to use, and no issues were noted.The surgeon was dissecting tissue at the time of the event, and the instrument performed as intended up until it broke.The customer used an irk, but it would not release the instrument due to the tip of the instrument being broken/snapped.There was no adverse effect to the grasped tissue and no unexpected tissue removal.There was no unexpected bleeding.There were no post-operative complications.It was confirmed there was no patient harm, and the patient was discharged from the hospital.
|