Catalog Number UNK_KIE
Device Problem Difficult to Remove (1528)
Patient Problems Discomfort (2330); Foreign Body In Patient (2687); Implant Pain (4561)
Event Date 03/31/2023
Event Type  Injury  
Event Description
As reported: "used timex extraction kit for proximal humeral nail removal.The nail would not come out using the usual conical extractor.The male extractor 4mm was used to try clear bone inside the nail and broke whilst trying to extract nail.The broken part remains in the nail.Unable to extract nail and had to leave the nail in patient.Nb: the male extractor would not usually be used in this situation however nothing else was working therefore it was attempted to force the nail out in other ways.The patient had already been on the table for approximately 2 hours trying to extract the nail before the male extractor broke." additional information: the patient wanted it removed as they had discomfort when raising their arms up, it was thought the nail was not buried enough however once opened, the top of the nail was firmly buried beneath bone.It was a planned removal due to patient discomfort and the surgeon is not planning to make another attempt to remove the nail.
 
Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned and no additional information was available.A device inspection was not possible since the affected device was not returned and no other evidences were provided for investigation.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN NAIL
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17019192
MDR Text Key316119578
Report Number0009610622-2023-00177
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberUNK_KIE
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/04/2023
Initial Date FDA Received05/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-