Model Number 74025771
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/04/2023
Event Type  malfunction  
Manufacturer Narrative
Complaint reference number: case (b)(4).
 
Event Description
It was reported that, during tka surgery, it was noticed that the sterile pack of one (1) jrny ii isrt xlpe dd rt sz 7-8 9mm was not sealed.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
Section h3, h6: the visual inspection revealed only an opened carton box and the implant was returned.No inner pouch was returned for evaluation.A review made by the quality engineering team revealed that the inner packaging material was not returned for evaluation and no pictures provided of the unsealed inner pouch, therefore, a thorough investigation could not be rendered.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According to the packaging sequence the sterile packaging should be sealed.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include damage during shipping and/or mishandling.Based on this investigation, the need for corrective action is not indicated.Should the inner packaging or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II ISRT XLPE DD RT SZ 7-8 9MM
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17022552
MDR Text Key316148682
Report Number1020279-2023-01200
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556098813
UDI-Public00885556098813
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113482
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74025771
Device Catalogue Number74025771
Device Lot Number19HM05406
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/04/2023
Initial Date FDA Received05/30/2023
Supplement Dates Manufacturer Received06/21/2023
Supplement Dates FDA Received06/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-