NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC GENERATION 2B SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5070S
Device Problems Break (1069); Corroded (1131); Physical Resistance/Sticking (4012)
Patient Problem Failure of Implant (1924)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation.The evaluation was performed by the london implant retrieval center (lirc).The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.
 
Event Description
An investigation report from an implant retrieval center was received and reported that the rod passed force testing.Additionally, it was noted during surface damage grading that there was galling and corrosion.It was reported that the rod was removed due to it being broken; there was no adverse patient or user impact associated with this report.No additional information is available.
 
Event Description
No additional information has been provided.
 
Manufacturer Narrative
The device was not returned to nuvasive for evaluation, however, the provided patient x-rays were reviewed and confirmed a solid section break.Without the return of the device, the root cause is unable to be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC GENERATION 2B SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr. ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key17032425
MDR Text Key316254424
Report Number3006179046-2023-00242
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00887517834027
UDI-Public887517834027
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMC2-5070S
Device Lot Number9030207AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/05/2023
Initial Date FDA Received05/31/2023
Supplement Dates Manufacturer Received08/23/2023
Supplement Dates FDA Received08/28/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/10/2017
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age8 YR
Patient SexMale
-
-