Model Number 10-80-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/02/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a s5 roller pump did not run and gave an error message.The issue occurred during procedure.There was no patient injury.
|
|
Manufacturer Narrative
|
H10: a livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue: roller pump displayed an error code associated to a fault in motor controller.In order to solve it, motor control and motor power amplifier boards were replaced and software flashed to updated version.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Complaints database analysis revealed that no similar event on this device occurred since its installation in 2020.Most likely pump stop can be related to a failure of the motor control due to an electronic failure, associated with the replaced boards.
|
|
Event Description
|
See initial report.
|
|