E1 - initial reporter facility name: (b)(6).E1 - initial reporter phone: (b)(6).Device evaluated by mfr.: epic vascular 8x80x120, was received for analysis.This epic device is recommended for use with a 0.035-inch guidewire as per epic instructions for use.During the product analysis a boston scientific 0.035-inch guidewire could not be inserted through the device due to damage to the inner sheath.The guidewire used by the customer was not returned for analysis.The device was returned with the stent partially deployed on the delivery system.To facilitate an examination of the inner sheath the investigator fully deployed the stent from the device.No issues or damage was noted with the stent.A visual and tactile examination identified twisting damage to the inner sheath of the device at approximately 75mm proximal of the tip.This type of damage is consistent with the tip of the device being twisted when handling the device, potentially when loading it on to the guidewire.
|
Reportable based on device analysis completed on (b)(6) 2023.It was reported that tracking difficulties were encountered.Vascular access was obtained via the right femoral artery.The 80% stenosed target lesion was located in the moderately tortuous and severely calcified left femoral artery.After a 6f non-boston scientific (bsc) guide catheter was placed and a 0.035 amplatz guidewire was crossed the lesion, an 8x80x120 epic vascular stent was advanced for treatment.However, the tip of the stent could not cross about 9cm into the end of the guidewire.The problem still occurred even after the physician replaced the guidewire.The device was removed, and the procedure was completed with another of the same device.There were no patient complications reported, and the patient status was stable.However, returned device analysis revealed a stent partial deployment.
|