It was reported calibration was successful with the two pressurewire x (pwx) wireless.However, equalization failed; the pressure signal dropped from time to time.The same issue occurred with the two pwx, during use.The transmitter light at the time of the issue is unknown.The distal tube for both pressurewires was separated from the transition tube, but still held together by the corewire.There were no adverse patient effects and no clinically significant delay during the procedure.No additional information was provided.No additional information was provided.
|
Visual analysis, functional testing, and electrical testing were performed on the returned device.The reported event of faulty pressure registration could not be confirmed due to device conditions.However, the reported material separation was confirmed.Functional testing was performed, but it failed to calibrate.An electrical test was performed which revealed that there was an open circuit on the proximal tube region.It was noted that the distal tube was separated from the transition tube, but still held together by the corewire.The investigation determined that the reported faulty pressure registration was likely due to the noted open circuit which could be caused by the noted separation on the distal tube.It is unknown what caused the separation.It may be possible that an excessive force was used during the insertion/withdrawal of the device, or there was an interaction between pressurewire with another device during the procedure which could lead to the noted separation; however, this could not be confirmed as it was based on operational circumstances.Based on the analysis and received information, the separation was still held together by the corewire, and there were no adverse patient effects.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|