The device was returned to olympus for evaluation and repair.During the evaluation, the reported issue could not be confirmed as no faults were detected.The investigation is still in progress; therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Correction to g2, adding other and china.Information was inadvertently not included on the initial medwatch.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.Based on the results of the investigation, and since no device malfunction was confirmed during evaluation, the definitive root cause of the reported issue could not be determined.The following information is stated in the instructions for use (ifu): ¿inspection of the endoscopic image¿ olympus will continue to monitor field performance for this device.
|