• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0702-XTW
Device Problems Product Quality Problem (1506); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/25/2023
Event Type  malfunction  
Event Description
This is filed to report torn material.It was reported this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 3.An xtw clip was inserted into the left atrium (la).The clip opened without issues.However, while closing the clip, it was observed one clip arm was not at the same angle to the shaft as the other clip arm.Therefore, the clip closed and replaced.Two clips were then implanted, reducing mr to a grade of 1-2.There were no adverse patient effects and no clinically significant delay in the procedure.The returned device analysis reported the clip introducer was torn.No additional information was provided.
 
Manufacturer Narrative
All available information was investigated, and the reported uneven clip arms was confirmed via returned device analysis.Additionally, a bent threaded stud and torn clip introducer were observed.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and the returned device analysis, the cause of the observed bent threaded stud was unable to be determined.The reported uneven clip arms was a cascading event of the observed bent threaded stud.The observed torn clip introducer was due to post-procedural handling of the device and are unrelated to the reported event.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17046149
MDR Text Key316385123
Report Number2135147-2023-02420
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/23/2024
Device Catalogue NumberCDS0702-XTW
Device Lot Number30124R1104
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2023
Initial Date FDA Received06/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
-
-