SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number CA25B0/400/000JP |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the product was observed.No adverse patient effects were reported by the customer.
|
|
Manufacturer Narrative
|
Other, other text: d4: expiration date and manufacture date are unknown, no information is available based on reported lot number.Date of event and udi number is unknown, no information has been provided to date.510k is blank device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Device evaluation: two devices were returned for investigation.Visual inspection revealed no abnormalities such as damage related to the reported event.A leak test was conducted on the breathing circuit and breathing bag, but no leaks were confirmed.It was confirmed that the product conformed to the standard.The reported events were not confirmed.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.This issue will continue to be monitored and further actions taken accordingly.
|
|