SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101329-NLJ |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3: month and year have been provided, day is unknown.D4: udi section is unknown, no information is available to date.D4: expiration date and h4: manufacture date are not available based on the reported lot number.G5: 510k is blank device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the product was observed.No adverse patient effects were reported by the customer.It was reported that no additional information is available.
|
|
Manufacturer Narrative
|
Device evaluation: one device and two photos were returned for investigation.Visual inspection occurred and found a damaged connector in the returned device.A functional leak test was conducted and the leaking issue was not present.The reported complaint is not confirmed.However, the damaged connector was found.Based on the analysis of the provided sample and photos, the most probable cause for the damage is improper handling of the product.No corrective actions are required.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
|