Catalog Number C37101329-NLJ
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2023
Event Type  malfunction  
Manufacturer Narrative
B3: month and year have been provided, day is unknown.D4: udi section is unknown, no information is available to date.D4: expiration date and h4: manufacture date are not available based on the reported lot number.G5: 510k is blank device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that during the pre-use check, leakage of air from the product was observed.No adverse patient effects were reported by the customer.It was reported that no additional information is available.
 
Manufacturer Narrative
Device evaluation: one device and two photos were returned for investigation.Visual inspection occurred and found a damaged connector in the returned device.A functional leak test was conducted and the leaking issue was not present.The reported complaint is not confirmed.However, the damaged connector was found.Based on the analysis of the provided sample and photos, the most probable cause for the damage is improper handling of the product.No corrective actions are required.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUIT
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17054700
MDR Text Key316476769
Report Number3012307300-2023-06012
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC37101329-NLJ
Device Lot Number4337191
Was Device Available for Evaluation? No
Date Returned to Manufacturer05/29/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/22/2023
Initial Date FDA Received06/02/2023
Supplement Dates Manufacturer Received07/28/2023
Supplement Dates FDA Received08/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-