• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. CAMERA HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. CAMERA HEAD Back to Search Results
Model Number MAJ-554
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/09/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, during preparation for a diagnostic procedure, the camera head had image problem.Subsequently, the intended procedure was completed with a similar device.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
The subject device has been returned to olympus for evaluation and the reported issue was confirmed.During evaluation, it was observed, glass was noted to be broken, the guide pun of the coupler was broken, and the fixing part of the coupler was loose.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the image issue was caused by a faulty head part.However, the definitive root cause of the faulty head part and broken glass could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAMERA HEAD
Type of Device
CAMERA HEAD
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17062055
MDR Text Key317638210
Report Number3002808148-2023-05609
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170136931
UDI-Public04953170136931
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K955404
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-554
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/09/2023
Initial Date FDA Received06/05/2023
Supplement Dates Manufacturer Received06/21/2023
Supplement Dates FDA Received07/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MODEL NUMBER: OTV-SC, NO SERIAL NUMBER
-
-