During device evaluation at olympus, it was found the complaint was not confirmed and no fluid invasion was found.Evaluation did find the front panel mouth was cracked, the top cover was rusted, the bottom and front side chassis was rusted, there was unknown residue on the rear panel, the socket slider switch was worn out which caused intermittent use of high intensity mode, the scope socket was corroded, and a non-olympus lamp life meter was reading over 500 hours.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
|