Model Number 71362295 |
Device Problems
Fracture (1260); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Complaint reference number: (b)(4).
|
|
Event Description
|
It was reported that, during a thr surgery, the rotating head of a ref ball jnt screwdriver shaft snapped off in the screw while inserting it.The broken piece was removed with pliers.The procedure was resumed, after a non-significant delay, with a s+n backup device.No other complications were reported due to this issue.
|
|
Manufacturer Narrative
|
The device was not returned for evaluation; therefore, a device analysis could not be performed.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for care, maintenance, cleaning and sterilization of smith & nephew orthopedics devices revealed that visually inspecting for damage or wear, including components in their disassembled state prior to re-assembly as well as ensuring components are re-assembled securely is indicated.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Manufacturer Narrative
|
Section h3, h6: the associated device was returned and evaluated.A visual inspection of the returned device reveals the spring, spring seat and screwdriver are missing from the device.These pieces were not returned.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|