• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306594
Device Problem Partial Blockage (1065)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2023
Event Type  malfunction  
Event Description
It was reported while using bd posiflush¿ normal saline syringe the plunger movement was difficult.There was no report of patient impact.The following information was provided by the initial reporter, translated from chinese to english: when the nurse seals the tube for the patient, the residual gas pressure in the prefilled catheter irrigator is large, which is difficult to discharge, and the liquid in the tube is wasted more, and the positive pressure joint of the patient can not be closely connected, which is easy to loose.
 
Manufacturer Narrative
H6: investigation summary, a device history record review was completed by our quality engineer team for provided material number 306594, and lot number 2006817.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17073253
MDR Text Key317037958
Report Number1911916-2023-00373
Device Sequence Number1
Product Code NGT
UDI-Device Identifier00382903065943
UDI-Public(01)00382903065943
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number306594
Device Lot Number2006817
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/12/2023
Initial Date FDA Received06/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-