It was reported, unable to advance the stent due to resistance.Resistance was felt when the stent was inserted into a microcatheter.The stent was therefore withdrawn and removed from the patient.The delivery wire tip was found to be broken after removal from the patient.The stent was not used and was replaced with another stent to continue the procedure.Subsequently, the procedure was successfully completed with no health damage to the patient.
|
The stent was returned loaded in the introducer.The distal tip of the pusher was found to be damaged.The stent was advanced into a new microcatheter for investigation and did not encounter resistance.The stent was able to deploy normally during the investigation.The proximal marker band od is specified at 0.0175¿ (+/-0.0003), and the returned pusher measured 0.0174¿.The introducer distal tip id is specified at 0.0215¿ (+0.0001/-0.0005), and the returned introducer measured 0.0214¿.The investigation of the returned stent system found the pusher distal tip damaged; however, no break condition was observed on the pusher distal tip as described in the reported event.Furthermore, the device was able to enter an in-house microcatheter, advance, and deploy without any resistance during functional testing.The physical evaluation of the device could not identify the conditions or circumstances that led to the pusher damage, but the damage is consistent with the device experiencing forces over specification.The microcatheter used in the procedure was not evaluated as a part of this evaluation, so the investigation could not determine if it had caused or contributed to the reported complaint.
|