G2: this event occurred in canada, see e1-e3.H3, h6: a medtronic representative visited the site to evaluate the equipment.Fluoro gain calibration and rad gain calibration were completed.Codes b01, c13, and d02 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
Medtronic received information regarding an imaging system being used during a spinal procedure.It was reported that the site encountered an issue during a case "last week" where after taking a spin, the image was completely black, and when transferred to picture archiving and communication service (pacs), they could see lines across the scan space.They attempted to completed a rad/gain cal but it failed at first.The second time they were able to complete the calibration, but imaging and navigation were aborted.There was no reported delay to the procedure.There was no reported impact to the patient.
|