Catalog Number 8065753057
Device Problems Failure to Deliver Energy (1211); Self-Activation or Keying (1557)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that an ophthalmic operating console failed multiple times and exhibited no phacoemulsification power.The touch screen displayed ultra-chop at bottom port when port is not was not in use.The procedure details and patient impact were not reported.Additional information has been requested but none received till date.
 
Manufacturer Narrative
The company representative was able to resolve the reported event by discovering that two handpieces (hps) were plugged into the console when it only should have one plugged in one at a time.This impacted ¿phaco power.¿ additionally, one of the handpieces (plugged in), was another kind of hp which is what prompted ¿ultra-chop¿ from the system.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed there were none relevant to this event, found.The root cause of the reported event is attributed to handpiece customer misuse and is unrelated to the functionality of the device itself.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM (ACTIVE SENTRY)
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17077133
MDR Text Key316663035
Report Number2028159-2023-00806
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657530571
UDI-Public00380657530571
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065753057
Device Lot Number143YC3
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/10/2023
Initial Date FDA Received06/07/2023
Supplement Dates Manufacturer Received09/05/2023
Supplement Dates FDA Received09/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-