Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi70000110, lot/serial #: unknown.H3 the manufacturer representative went to the site to test the imaging system.It was found that the door was not fully opening and when closed the "door open" sign was displayed on the pendant.After removing some covers (inner door covers and inner gantry covers) it was indeed identified that the dingus was misaligned.The dingus were repositioned manually and the rotor was manually brought to the proper position for opening and closing the door.Door winch cable was examined and no issues were found.Several door openings and closings took place without the covers and with the covers back in place.System functions have been successfully tested.The issue has been resolved.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|