According to the product experience report, there was no sample to be returned.Additionally, no photographic or visual evidence of any kind was provided which would have allowed for the allegation to be reviewed.Without such evidence, this complaint could not be confirmed and determining a definite root cause and corrective action is not possible.If the sample is returned at a future date, a follow-up report will be submitted.
|
Argon 2.6 fr dl distal extension tube of the red lumen cracked.The line was inserted on (b)(6) 2023, cracked, removed, and replaced on (b)(6) 2023.¿what was the original intended procedure? : parenteral nutrition, iv fluids, intermittent iv antibiotics¿ picc insertion.
|