It was reported to vyaire medical that the etco2 (end-tidal carbon dioxide) level on 2079796-001 - medisorb ex, pkg of 6 pcs, showed higher than normal during usage.The customer also mentioned that they are unable to identify how long it was used.The customer confirmed that there was no patient harm or injury associated with the reported event.
|
Device evaluation: g3, g6, h2, h3, h6 and h10 production records showed no deviations from the specification of the device during manufacture.All units packed passed a 100% leak test and visual inspection.A review of the batch documentation under lot l01a-01340 found no issues that could be linked to damage.The returned sample has been visually inspected, and damage found at the canister¿s bottom has been identified, which would indicate that the most likely cause is that the canister was subjected to a significant force during transportation after the 100% in-line leak and visual inspection check in the manufacturing line.The device returned was in very poor condition, severely damaged, and damp, causing the material contained in it to fail all performance tests.Material retention samples were tested for caco3, usp co2 absorption, and moisture absorption, and both passed the tests within spec.The caco3 result shows that although the soda lime has been used, there is still capacity for the device to absorb co2 (exhaustion results typically range from 60 to 80%).Due to the device being decontaminated with solution (externally and internally), this affected the moisture and the performance of the soda lime, and mpl cannot ascertain if the material from this device was ok before being decontaminated.Retain samples have been tested and show that the material is still performing as intended if used and stored correctly.
|