SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101307J |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the product was observed.No adverse patient effects were reported by the customer.It was reported that no further information is available.
|
|
Manufacturer Narrative
|
Month and year of event have been provided, day is unknown.510k is unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
One device and three photos were used for evaluation.Visual inspection and functional testing occurred and found embedded resin on the corrugated tube and a leak was present.The complaint is confirmed.Based on the analysis conducted in the sample provided, this problem has been caused during the extrusion process.This issue will continue to be monitored and further actions taken accordingly.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
|