Model Number X SERIES |
Device Problem
Smoking (1585)
|
Patient Problem
Chest Pain (1776)
|
Event Date 05/18/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that while attempting to monitor a 17-year-old female patient, they observed a very small amount of smoke coming from the attached ecg electrodes.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
The device and ecg monitoring leads were returned to zoll medical united kingdom for evaluation.The customer's report was not replicated or confirmed.The device was put through extensive testing including electrical safety testing using the customer's returned leads without duplicating the report.An internal inspection of the device found no discrepancies.The ecg monitoring leads were discarded as a precaution.The device was recertified and returned to the customer.It is important to note that ecg leads are not a current pathway and do not deliver energy.No trend is associated with reports of this type.
|
|