Model Number WA22302D
Device Problems Failure to Cut (2587); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2023
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
It was reported when performing a therapeutic transurethral plasma resection of the prostate the device was used to cut the tissue, it was found that the tissue could not be cut off when it reached to the root of the tissue, and the wire at the distal end was burnt black.The device was replaced and the intended procedure was completed with a similar device.There were no reports of patient harm or delay.No user injury reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the burnt wire was wear and tear.The loop at the distal end of the electrode wears out during use and may break, burn or melt.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP
Type of Device
RESECTION ELECTRODE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17084106
MDR Text Key317509743
Report Number9610773-2023-01583
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761051665
UDI-Public14042761051665
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K100275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA22302D
Device Catalogue NumberWA22302D
Device Lot Number1000102788
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/09/2023
Initial Date FDA Received06/07/2023
Supplement Dates Manufacturer Received06/27/2023
Supplement Dates FDA Received07/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-