Model Number 777626 |
Device Problems
Inadequate Instructions for Healthcare Professional (1319); Wrong Label (4073)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/17/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that the contents inside does not match the packaging.The packaging was marked 777626 but the contents inside were 777426.
|
|
Event Description
|
It was reported that the contents inside does not match the packaging.The packaging was marked 777626 but the contents inside were 777426.
|
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Correction: f,h.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Manufacturer Narrative
|
The reported event was confirmed as packaging related.A potential root cause for this failure mode could be ¿incorrect component".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not required as a review of the label could not have prevented the reported event.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
|
Event Description
|
It was reported that the contents inside does not match the packaging.The packaging was marked 777626 but the contents inside were 777426.
|
|