Model Number 286750100
Device Problem Calibration Problem (2890)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the devices were received as blind units on april 24, 2023.Upon product inspection on may 24, 2023, it was discovered that the items were out of calibration.This report involves one viper prime torque handle.This is report 5 of 9 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.E3: reporter is a j&j employee.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there were no cosmetic defects observed on the surface of the device.A dimensional inspection was not performed as it is not applicable to the complaint condition.A functional test was performed.Measured torque values were found to be outside the specified range.The device was testing high.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the viper prime torque handle, p/n: 286750100, lot: gb116809 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a review of the receiving inspection (ri) for viper prime torque handle was conducted identifying that lot number gb116809 was released in one batch.Batch 1: lot qty of 110 units were released on 29 jan 2019 with no discrepancies supplier: gauthier biomedical, inc.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER PRIME TORQUE HANDLE
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key17086961
MDR Text Key316807311
Report Number1526439-2023-01024
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10705034507750
UDI-Public(01)10705034507750
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number286750100
Device Catalogue Number286750100
Device Lot NumberGB116809
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2023
Initial Date FDA Received06/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-