• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number AZURION 7 M20
Device Problem Break (1069)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/30/2023
Event Type  malfunction  
Event Description
It has been reported to philips that the c-arm broke resulting in yanking a patient's shoulder.Also the arm support board was broken.The patient outcome is unknown and it is unknown if there was actual harm to the patient.Philips is collecting additional information.We are conservatively reporting this event as an injury while the investigation is ongoing.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was in clinical use for treatment when the issue occurred and the patient's shoulder was pulled back but the patient was unharmed, and no injury occurred to the patient and the procedure was completed as planned.The customer reported that the c-arm cable caught the arm board on rotation and broke the arm board.The customer ordered and replaced the arm support board.After replacement of the arm support board, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key17096643
MDR Text Key316848805
Report Number3003768277-2023-03493
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00884838085268
UDI-Public00884838085268
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K163715
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAZURION 7 M20
Device Catalogue Number722079
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/30/2023
Initial Date FDA Received06/09/2023
Supplement Dates Manufacturer Received09/15/2023
Supplement Dates FDA Received10/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-