Catalog Number H938724E
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the outer packaging of an em2400 valve set was not correctly sealed and probably no longer sterile.The unsealed/opened packaging was observed before patient use of the device.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter first name - (b)(6) e1: initial reporter address: (b)6) should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: one unused device was received for evaluation.An unaided visual inspection was performed and observed that the primary packaging seal was not completely sealed through at the top side area; an opened section approximately 10 mm in length was observed at top right side.Magnified inspection found no visible contamination of product or packaging.The reported condition was verified.The cause of the condition could not be determined; however, a possible cause may be due to impact damage sustained during transport or due to variation during the manufacturing packaging process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17097637
MDR Text Key316858526
Report Number1416980-2023-02817
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412475776
UDI-Public(01)00085412475776
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 07/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938724E
Device Lot Number60372883
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/16/2023
Initial Date FDA Received06/09/2023
Supplement Dates Manufacturer Received07/12/2023
Supplement Dates FDA Received07/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-